High Flow rate – recirculate calibration system | ||||
Capacity : | 4.6 ~ 2000 | Am3/h actual flow rate | ||
4.6 ~ 20000 | Nm3/h standard flow rate | |||
Op. Pressure: | 1~ 10 bar | Absolute pressure | ||
Smallest uncertainty | 0.30% | |||
Apply for:: | ||||
Turbine , Rotary , Ultrasonic , Vortex , Thermal(thermal mass) , wheel | ||||
Positive displacement, Differential pressure…etc., types flow meter. |
- With a whole year’s construction, we had built up a central control calibration system, which will automatic holding steady pressure/temperature while calibration. It can operate at pressure 1 to 10 BarA, and holding the temperature steady within ±0.1°C during the calibration.
- The actual flow rate is from 4.6 to 2000 Am3/h (Actual volume flow rate at working pressure and temperature).
- The standard flow rate is from 4.6 to 20000 Nm3/h (standard volume flow rate is as the volume based on 0°C,1atm).
- Applicable pipeline is from 1 to 8 inch ( DN25 ~ DN200 ),
- This system was ISO/IEC 17025 accredited by TAF on 2009, and by iLAC-MRA , this accreditation was equal to be accredited by DAkks/JAB/IAJapen/UKAS/A2LA/NVLAP/IAS, and will be accepted by 73 accreditation organizations among countries all over the world.
- The smallest uncertainty is excellently 0.30%.