Capacity: | 0.025 ~ 1667 | LPM | ||||
Op. pressure: | regular | 0~3kg/cm3 | ||||
Smallest Uncertainty | 0.50% | |||||
Apply for: | ||||||
Electro-magnetic , Turbine , Rotary , Ultrasonic , Vortex , wheel , | ||||||
Positive Displacement, Rota(Variable Area) , Differential Pressure , ..etc. |
Head tank calibrate system, literately, it has a 10 meters high water tank which provide a steady water head, and it is resulting a very steady water flow rate when calibrating. it’s also equipped temperature steady controller.
This system was composed by two systems:
1)Comparative system-compare with our standard flow meters, flow rate is form 5 to 1667 LPM.
headtank
2)Weighting system-it’s a primary calibrate system by weight-flow rate converting, flow rate is from 0.025 to 5 LPM. The smallest uncertainty is 0.5%.
This system was accredited by TAF recently.